Set the ALLOW_IGNORE_QC_FAILURES Preference

Use this activity to set the ALLOW_IGNORE_QC_FAILURES application preference to allow users to ignore a composite QC failure.

The Scope for this preference is usually set at user level.

Activity Steps

  1. Access the CCPREF—Preference application from the main menu or quick launch to display the CCPREF—Search Preference screen.
  2. Enter the Preference NameALLOW_IGNORE_QC_FAILURES
  3. Click Submit.
  4. Open the preference with the lowest level of scope to display the CCPREF—Preference Detail screen.
  5. If Is System is checked, then:
    1. Uncheck Is System.
    2. Click Submit.
  6. Ensure that Setting Type is set to BOOLEAN.
  7. To allow a user to ignore a composite QC failure, check Setting Value.

    In the case that a result fails scheme limit assessment, standard assessment, precision assessment or primary specification assessment, such that, the Composite QC Status is set to Failure thereby stopping validation so the the Workflow Status cannot proceed to Completed, the user has the option to ignore the composite QC status, upon which the Composite QC Status is set to Ignored Failure, making result validation possible so that the Workflow Status proceeds to Completed.

  8. To prohibit a user from ignoring a composite QC failure, uncheck Setting Value.

    If a result fails scheme limit assessment, standard assessment, precision assessment or primary specification assessment, the user cannot ignore the composite QC failure.

  9. Set the Preference Type to PREFERENCE.
  10. Ensure Is Active is checked.
  11. Click Submit.
  12. If Is System was initially checked, then:
    1. Check Is System.
    2. Click Submit.