Forcing the Reassessment of Results in a Workbook Session
Overview
Once a result is entered and the system evaluates the composite QC status, if the sample scheme analyte's Composite QC Status is set to Failure, then scheme limits, accuracy or precision limits and/or primary specification limits have failed. This can be overcome by either:
- Overwriting the result with a value that is within limits, derived somehow from a correct analysis, using Overwriting Result in a Workbook Session.
- Requesting a repeat analysis so that another sample portion is used to derive the result using Requesting a Repeat Analysis in a Workbook Session.
- Ignoring the composite QC failure using Ignoring a Composite QC Failure in a Workbook Session, but only where there is a valid reason to do so.
- Changing the assessment limits and then perform a reassessment using the this process, but only where there is a valid reason to do so.
Process
Force the reassessment of results in workbook cells
QC History Observation Update
Any update to a QC sample scheme analyte's Workflow Status is persisted to the associated QC history observation.
- Understanding System Processes that follow Result Entry
- Assessing Results against Detection Limits
- Assessing Results against Scheme or Surrogate Limits
- Assessing Results against Accuracy Limits
- Assessing Results against Precision Limits
- Assessing Results against Product and Specification Limits
- Appendix—Cascade Engine