Maintain job schemes and analytes using default selection

  1. Open a registered job to display the PROFILE—JobEdit Form.
  2. Select the Schemes tab.

To add a scheme to the job:

  1. Click New to add a scheme at the end of the job, or click Insert to insert a scheme above the currently selected row, to open the PROFILE—ProfJobSchemeAdd Form. The second column on display, after the Scheme column, is defined by the SchemeLVWDisplayColumn PROFILE program setting.
  2. Where the ProfJobSchemeAddNew.SCR script file is not located in the folder defined by the ConfigDirectory GLOBAL setting, then the list-view panel is loaded with all schemes.
  3. Where the script file is located, but the query event defined by the QueryEvent PROFILE program setting does not exist within the script file, then no schemes are returned to the list-view panel.
  4. Where the script file is located and contains the query event, then click Query to display the matching schemes in the list-view panel.
  5. Select the Scheme type to display schemes of that type.
  6. To add all analytes as a scheme is added in the next step, check the Add all Analytes check box. To not add analytes, but to selectively add them after a scheme is added, uncheck the Add all Analytes check box.
  7. Where the add-all-analytes is activated, select the scheme and click OK to add the scheme with all its analytes to the job, or click Cancel to not proceed.
  8. Where the add-all-analytes is not activated, select the scheme and click OK to add the scheme to the job and open the PROFILE—ProfJobAnalyteList Form. Use the >, >>,<< and < buttons to assign Available analytes to the Assigned analytes and click OK to add the analytes to the job, or click Cancel to not proceed.

To review a scheme in a job:

  1. Double-click a Scheme to open the PROFILE—ProfJobSchemeEditANA Form, PROFILE—ProfJobSchemeEditPRP Form or PROFILE—ProfJobSchemeEditMISHDR Form, where the scheme's Type is ANA, PRP, or MIS or HDR, respectively.

To update a scheme in a job:

  1. Double-click a Scheme to open the SCHEME—ProfJobSchemeEditANA Form, SCHEME—ProfJobSchemeEditPRP Form or SCHEME—ProfJobSchemeEditMISHDR Form, where the scheme's Type is ANA, PRP, or MIS or HDR, respectively.
  2. Click Save to close the form and save the changes, or click Close to close the form.
    Where the ProfJobUDF.VBS script file is located in the folder defined by the ConfigDirectory GLOBAL setting, and the file contains the SchemeUpdated subroutine, then the subroutine is run.

To reorder the schemes of the job:

  1. Select any scheme in the Scheme column.
  2. Click Move to open the PROFILE—ProfJobSeqMove Form. Use the Up arrow and Down arrow buttons to reorder the schemes within the job. Click OK to close the form and save the reorder, or click Cancel to close the form.

To reorder the analytes within a job scheme:

  1. Select analyte in the Analyte column for a scheme.
  2. Click Move to open the PROFILE—ProfJobAnalyteList Form. Use the >, >>,<< and < buttons to assign Available analytes to the Assigned analytes. Use the Up arrow and Down arrow buttons to reorder the analytes within the scheme of the job. Click OK to close the form and save the reorder, or click Cancel to close the form.

To enter analysis details for a PRP (Preparation) or ANA (Analytical) scheme in a job:

  1. Enter the Page size. For a scheme with a Type of ANA, these fields are located on the General tab.

To enter duplicate details for a PRP (Preparation) scheme in a job:

  1. Enter the Number, Blocksize, Offset and Position.

To enter quality control details for an ANA (Analytical) scheme in a job:

  1. Select the General tab.
  2. Blank placement:
    1. Enter the Number blanks.
    2. Select the Files tab and click ... next to Blank file contents to open the SCHEME—QCFile Form, and use the > and >> buttons to assign Available QC types to the Assigned QC types to add blank sample types for the required Number blanks (refer to Maintaining Quality Control Types). Alternatively, type in the blank sample types and click fx to confirm that they are valid ones. The same blank material can be assigned a multiple number of times to create the required number of blank placements.
    3. Select the General tab and enter the Blank blocksize, select the Blank calculation, and enter the Blank position(s).
  3. Standard placement:
    1. Enter the Number qc standards.
    2. Select the Files tab and click ... to open the SCHEME—QCFile Form, and use the > and >> buttons to assign Available QC types to the Assigned QC types to add QC standard sample types for the required Number qc standards (refer to Maintaining Quality Control Types). Alternatively, type in the QC standard sample types and click fx to confirm that they are valid ones. The same QC standard material can be assigned a multiple number of times to create the required number of QC standard placements.
    3. Select the General tab and enter the QC standard blocksize, select the QC standard calculation, and enter the QC standard position(s).
  4. Spike placement:
    1. Enter the Number qc spikes.
    2. Select the Files tab and click ... to open the SCHEME—QCFile Form, and use the > and >> buttons to assign Available QC types to the Assigned QC types to add QC spike sample types for the required Number qc spikes (refer to Maintaining Quality Control Types). The same QC spike material can be assigned a multiple number of times to create the required number of QC spike placements.
    3. Select the General tab and enter the QC spike blocksize, select the QC spike calculation and QC spike selection, and enter the QC spike position(s).
  5. Replicate placement:
    1. Enter the Number replicates.
    2. Enter the Replicate blocksize, select the Replicate calculation and Replicate selection, and enter the Replicate position(s). Example: To create a replicate of each sample and place each replicate sample immediately after its original unknown sample, set Number replicates to 1, Replicate blocksize to 1, and Replicate position(s) to either IMM or END, as this is equivalent in this scenario.
  6. Enter the Number rereads.

  7. Select the Files tab and select a Link scheme code, if required.

To enter invoicing details for scheme in a job, where the scheme's Price type is SCH, HLY or SMP:

  1. Enter the Price code or fixed Price. For a scheme with a Type of ANA, these fields are located on the General tab.
  2. Where the Price type is HLY, then enter the Number hours so that a price is determinable.
  3. To tie the scheme in with procedural pricing, select the Procedure code.

To enter custom details for a scheme in a job:

  1. Where the ProfJobSchemeUser.SCR screen file exists, click User Screen in the main menu to open the custom form to update the custom details, as required.

To delete a scheme from a job:

  1. Select the scheme in the Scheme column and click Delete.

To review an analyte in an ANA (Analytical) scheme in a job:

  1. Double-click an Analyte in a scheme that has a Type of ANA to open the PROFILE—ProcRegSchemeAnalyteEdit Form.

To update an analyte in an ANA (Analytical) scheme in a job:

  1. Double-click an Analyte in a scheme that has a Type of ANA to open the PROFILE—ProcRegSchemeAnalyteEdit Form.
  2. Click Save to close the form and save the changes, or click Close to close the form.

To update an analyte in an ANA (Analytical) scheme in a job:

  1. Select the Analysis unit code.
  2. Enter the Analysis detection limit and Analysis upper limit.
  3. Select a Reporting unit code and Rounding table code.
  4. Enter the Reporting detection limit and Reporting upper limit.
  5. Enter Lower limit 1, Lower limit 2, Upper limit 1 and Upper limit 2, as required.
  6. Click Save to persist the changes, or click Close to not proceed.

To enter custom details for an analyte in an ANA (Analytical) scheme in a job:

  1. Where the ProfJobSchemeAnalyteUser.SCR screen file exists, click User Screen in the main menu to open the custom form to update the custom details, as required.

To add schemes and analytes to a job using a procedure:

  1. Click New procedure to open the PROFILE—ProcedureAdd Form. The ProcedureLVWDisplayColumn PROFILE program settings defines the second column of the grid after the Procedure code column.
  2. Where the ProfJobProcedureAddNew.SCR script file exists, to return procedures to the Procedure grid based on criteria:
    1. The query form contained in the script displays at the bottom of the screen.
    2. Enter the criteria and click Query to run the event defined by the QueryEvent.
  3. Where the script file does not exist, then all procedures are returned.
  4. Select a Procedure code and click OK to close the form and add the delta schemes and analytes to the job, or click Cancel to close the form and not proceed.
  5. Click Save to persist the changes to the job.

To delete an analyte from a scheme in a job:

  1. Select the analyte in the Analyte column and click Delete.